9. Press IENTER] to confirm your calibration. The screen will momentarily display"Do CALIBRATION SAVED" as shown below.NOTE: If you wish to abort calibration before you have pressed[ENTER, you may press[MODE] to return to the calibrate menu without saving the new calibration value. You may also press[NEXT] to select a different parameter(any change made will not be saved) 10. Press[MoDE] to return to the Main mode. The instrument is now calibrated and ready to measure dissolved oxygen and temperature. See 1 Main Mode. Making Measurements.Barometer.- The YSI Model 5100 has an internal barometer for pressure compensation during AUTO Dissolved oxygen Calibration. This barometer only needs to be calibrated when it is no longer reading the correct barometric pressure. If the 5100 is kept at a fairly constant ambient temperature(tl0 C), the barometer calibration should be accurate for approximately 30 days. The Model 5000 does not contain a barometer, therefore, the current barometric pressure must be entered before an AUTO Cal is performed. The pressure value displayed is the setting that was entered and stored during the previous calibration. From the calibration menu press the IDO CAL] soft-key, the[NEXT] soft-key until the barometric pressure is flashing and"Barometer" appears in the top right corner of the display as follows: